HR-L8013 Particulate Filtration Efficiency Tester is our newly developed automatic filter testing device. It is suitable for the efficiency and resistance test of 0.1%-99.99% filter material. The structure of device is designed compactly and lightweight. HR-L8013 Particulate Filtration Efficiency Tester only needs power supply and compressed air (no water, no oil) for running,which is widely used in laboratory, factory, etc.
Main Features:
1. Auto calculation Fabric air resistance and Textile air resistance.
2. Two remote Airborne particle counters are used simultaneously to detect the Aerosol concentration upstream and downstream, and make sure the sampling datas are authenticity and accuracy.
3. Atomizer is imported from Korea ART+ company, stable concentration, stainless steel anti-corrosion material.
4. 10 inch color touch screen, easy-to-use.
5. High-quality blower, auto-control flow rate.
6. It support remote maintenance, The system automatically calculates the test results and printout the test result.
Main Parts:
1. Atomizer
It is used for the test which need high concentrations of aerosols. It used 0.3um PLS standard particles. More than 70% particles , the size of it ≥0.3um, and it can generate steady and sustainable particles.
2. Airborne Particle Counter
It is used for test particle concentration up and down the test material. And get the test result.
3. Differential pressure sensor
It is used for test differential pressure (resistance) for material in test flow rate.
4. AL Fixture
It is used for press and seal the test material.
5. High-quality blower
Mainly used to provide test flow of test material, auto-control flow rate.
6. Electrostatic neutralizer
Remove charge in Nacl particles and make the test result more accurate.
HR-L8013 Particulate Filtration Efficiency Tester
Test efficiency range
0.1%-99.99%@≥0.3μm
Test area
100cm²
Resistance range
0-1000pa, accuracy 0.1%
Test flow
15-100L/min(adjustable)
Aerosol
Nacl, DEHS
Particle Counter flow rate
2.83L/min ±5%
Channels of Particle Counter
0.3um, 0.5um, 1.0um, 2.5um, 5.0um, 10.0um,
Pump of Particle Counter
Japanese pump, More than 5 years service life
Storage
100,000 group data and It can USB export data
Display
10 inch large color touch screen
Printer
Built-in Printer
Power supply
AC 220V/50Hz
Data export
USB
Temperature and Humidity
Temperate:0-40℃ Humidity:0-70%
Accessories
Printer paper, User Manual, Test report,
Factory Certificate, 200mLNacl,200mLDEHS
Dimension
L800x W600x H1450(mm)
Material
Steel plate for casing
Weight
120Kg
Warranty
1 Year