Principle of Operation
A known volume of air is drawn by vacuum through a slit orifice that accelerates the sampled air and its contaminants. Located below the slit and in the path of airflow is an exposed agar plate that rotates on a turntable at a selected rate of speed. The contaminants, because of their higher mass become impacted on the agar surface, while the rest of the air mass flows around the plate and exits the air sampler. After incubation of the agar plate, colonies can be counted and identified. A particle distribution grid can also be used to help determine when a contamination event occurred.
Specification of FSC-Ⅱ Biological Air Sampler:
Flow rate: 50L/min
Sample Exposure time(time for one revolution of a 150mm Petri dish): 1-99min
Petri dish size: Ф90×15mm
Sample volume: 50L-5000L
Vacuum source: built-in oil free vacuum pump
Dimension: 28.5×11.5×28.5 cm
Electrical requirements: 220v 50/60Hz
Weight: 8.5kg